Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
9743611 | Analytica Chimica Acta | 2005 | 7 Pages |
Abstract
A simple flow injection chemiluminometric method for the determination of tetracycline (TC), chlortetracycline (CTC) or oxytetracycline (OTC) has been developed. The method is based on the chemiluminescence (CL) produced the reduction by these tetracyclines of tris(2,2â²-bipyridyl)ruthenium(III). The latter is obtained by oxidation of tris(2,2â²-bipyridyl)ruthenium(II) by potassium permanganate in dilute nitric acid. A standard or sample solution is injected into the ruthenium(II) stream either before or after it has merged with the potassium permanganate stream. The CL intensity is increased by the presence of manganese(II). Under the optimum conditions, calibration graphs were obtained for 5 Ã 10â5 to 5 Ã 10â4 mol lâ1 of each tetracycline. The detection limits (s/n = 3) were 2.0 Ã 10â6, 1.9 Ã 10â6 and 1.0 Ã 10â6 mol lâ1, respectively, and mid-calibration range standard deviations were not greater than 3.7% (n = 10). The method was successfully applied to the determination of these drugs in pharmaceutical formulations with a sample throughput of 50 hâ1.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Alan Townshend, Wirat Ruengsitagoon, Chalermporn Thongpoon, Saisunee Liawruangrath,