| Article ID | Journal | Published Year | Pages | File Type |
|---|---|---|---|---|
| 9743646 | Analytica Chimica Acta | 2005 | 12 Pages |
Abstract
Microwave-assisted extraction (MAE) has been evaluated as an alternative to dialysis for extraction of some water-borne hydrophobic contaminants sampled by semi-permeable membrane devices (SPMDs). Seven organochlorine pesticides (OCPs), 11 polychlorinated biphenyls (PCBs) and 13 polycyclic aromatic hydrocarbons (PAHs) were accumulated in SPMDs at nanogram levels and extracted with three 3-min irradiation cycles with 33Â mL of a solvent mixture hexane-water (10:1,v/v) in each cycle. The developed MAE method gave for all analytes investigated statistically comparable extraction yields with those found by dialysis carried out with a total volume of 250Â mL hexane for 48Â h at room temperature. The recoveries of all the targeted contaminants were in the range of 65-105% with variation coefficients not exceeding 19%. The applicability of the MAE extraction was tested in field SPMDs samples deployed for 15 days in a sewage-treatment process. Our results show that MAE provides a remarkable reduction of time and solvent volume when used as an extraction method in the analysis of SPMDs.
Keywords
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
V. Yusà , A. Pastor, M. de la Guardia,