Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1105784 | Transfusion Clinique et Biologique | 2006 | 5 Pages |
Abstract
Nucleic acid testing is carried out in several steps: plasma pooling, nucleic acid extraction, amplification and detection. The target values of temperature, volume and incubation time are mentioned in the initial protocol without giving their limit values. The objective of this work was to determine the range of values in which the test remains positives. So we have tested: 1) the temperature and incubation time of plasma in lysis buffer (37 °Câ
屉
3 et 30Â minâ
屉
10); 2) the influence of volume and incubation time of silica (50 μlâ
屉
10 et 10Â minâ
屉
5); 3) the variation of the eluate volume after nucleic extraction. We have also studied the influence of the internal control volume variation. For each parameter the assays were performed at sensitivity limit of the technique and repeated several times. Our results showed that: 1) for all parameters evaluated the tests remain positive within a wide range of values; 2) It is not necessary to set up a sophisticated measurement process; 3) the technique is robust.