Article ID Journal Published Year Pages File Type
4495447 Journal of Northeast Agricultural University (English Edition) 2012 7 Pages PDF
Abstract

Reversed-phase high-performance liquid chromatography (RP-HPLC) method with C4 column and C18 column for analyzing β-lactoglobulin and α-lactalbumin in bovine milk was developed and the performance and characteristic of two columns were compared. Shiseido Proteonavi C4 column (250 mm×4.6 mm×5 μm) and Shiseido CAPCELL PAK SG 300 C18 column (250 mm×4.6 mm×5 μm) were used in the experiment. Phase A was composed of 0.1% (V/V) trifluoroacetic acid (TFA) in ultrapure water and Phase B (organic phase) was composed of 0.1% TFA in acetonitrile. Gradient elution was taken. Flow rate was 1 mL min−1. The detection wavelength was 215 nm. The injection volume was 20 μL and the column temperature was 30°C. The results showed that linear relationship was good and recovery of α-lactalbumin and β-lactoglobulin was 86.12%–104.38%, C18 column had stronger ability to resist acid and more stable, and the method with C4 column had excellent sensitivities and good separation.

Related Topics
Life Sciences Agricultural and Biological Sciences Agricultural and Biological Sciences (General)