Abstract
A 2 μohm drift was observed in the resistance measurements on a 24-h period. Two calibration methods were studied. The first calibration method was developed to calibrate the conductivity apparatus using power resistors and we were able to determine that a systematic error of 60 μohm was present. A second calibration method was then used to measure the electrical conductivity of two Poco graphite samples. Using our apparatus, the electrical conductivity of AXF-5Q and DFP-2 Poco graphite samples were both (60-210 μohm cm) lower than their reported statistical value of 1470 and 1500 μohm cm, respectively.
Keywords
Related Topics
Physical Sciences and Engineering
Chemistry
Electrochemistry
Authors
Nicolas Cunningham, Michel Lefèvre, Guy Lebrun, Jean-Pol Dodelet,