'Aussie Normals': an a priori study to develop reference intervals in a healthy Australian population using the Beckman Coulter LH 750 Haematology Analyser as candidates for harmonised values
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی پزشکی قانونی
پیش نمایش صفحه اول مقاله
'Aussie Normals': an a priori study to develop reference intervals in a healthy Australian population using the Beckman Coulter LH 750 Haematology Analyser as candidates for harmonised values
چکیده انگلیسی
Reference limits or intervals are important benchmarks or tools that help the clinician to distinguish between a result that is most likely to lie within a 'healthy' or diseased category. It has been suggested that a review of haematology reference intervals is long overdue. In this study we report on our findings for analytes routinely measured in a complete blood count (CBC) performed on the Beckman Coulter LH 750 analyser and an additional comparative study using the Beckman Coulter LH 750, the Sysmex XN and Abbott Sapphire. The results from the comparative study indicate that bias would not prevent harmonisation of reference intervals for these common haematology parameters. The results offered by the Aussie Normals study represent good candidates as the basis for harmonisation reference intervals.