A parametric analysis of fixation post shape in tibial knee prostheses
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
پیش نمایش صفحه اول مقاله
A parametric analysis of fixation post shape in tibial knee prostheses
چکیده انگلیسی
In terms of interface forces, a tapered post decreased post shear while slightly increasing post compression compared to a cylindrical post; a post of elliptical cross-section increased post shear and decreased post compression. In terms of cancellous bone stress, tapered and elliptical posts both relieved compression compared to a cylindrical post, while a tapered post also produced increased peripheral stress. The inclusion of medial and lateral pegs in addition to a central fixation post caused localized stress shielding in the periphery of the pegs. In general, all implant models caused a reduction of cancellous bone stress plus high compression beneath the central fixation posts.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Medical Engineering & Physics - Volume 27, Issue 2, March 2005, Pages 123-134