Reductions in MRI-induced heating at 1.5 T resulting from a simple design change to coiled wires were investigated. MRI-induced heating was assessed for two different coiled wire forms (length, 26 cm): (1) multi-filar coiled wire form and (2) multi-filar coiled wire form having a different coiled pitch, providing an air gap spacing between adjacent five-filar coil loops. Each wire had an electrode and was insulated to create a lead, similar to that which would be used for a medical implant. The wire forms were placed in a gelled-saline-filled head/torso phantom and imaged at 1.5 T [whole-body average specific absorption rate (SAR), 1.79 W/kg]. Fluoroptic thermometry probes were used to measure temperatures at the distal ends of the wires. The experiments demonstrated a substantial reduction in MRI-induced heating for the modified wire compared to the unmodified wire (i.e., 10.5°C difference observed in one experiment and 26°C difference in another). These findings have important implications for MRI-induced heating of leads used for medical implants.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Magnetic Resonance Imaging - Volume 23, Issue 8, October 2005, Pages 887-891
Journal: Magnetic Resonance Imaging - Volume 23, Issue 8, October 2005, Pages 887-891
نویسندگان
Robert W. Gray, W. Timothy Bibens, Frank G. Shellock,