کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1170742 | 960689 | 2007 | 9 صفحه PDF | دانلود رایگان |
A new technique for the determination of suspended particulate trace metals (P–metals >0.2 μm), such as Co, Ni, Cu, Zn, Cd and Pb, in open ocean seawater has been developed by using microwave digestion coupled with flow injection inductively coupled plasma mass spectrometry (FI-ICP-MS). Suspended particulate matter (SPM) was collected from 500 mL of seawater on a Nuclepore filter (0.2 μm) using a closed filtration system. Both the SPM and filter were completely dissolved by microwave digestion. Reagents for the digestion were evaporated using a clean evaporation system, and the metals were redissolved in 0.8 M HNO3. The solution was diluted with buffer solution to give pH 5.0 and the metals were determined by FI-ICP-MS using a chelating adsorbent of 8-hydroxyquinoline immobilized on fluorinated metal alkoxide glass (MAF-8HQ). The procedure blanks with a filter were found to be 0.048 ± 0.008, 10.3 ± 0.3, 0.27 ± 0.05, 3.3 ± 1.8, 0.02 ± 0.03 and 0.85 ± 0.09 ng L−1 for Co, Ni, Cu, Zn, Cd and Pb, respectively (n = 14). Detection limits defined as 3 times the standard deviation of the blanks were 0.023, 0.90, 0.14, 5.3, 0.078 and 0.28 ng L−1 for Co, Ni, Cu, Zn, Cd and Pb, respectively. Accuracy was evaluated using certified reference materials of chlorella (NES CRM No. 3) and marine sediment (HISS-1). The method was applied to the determination of vertical distributions for P–Co, Ni, Cu, Zn, Cd and Pb in the Western North Pacific.
Journal: Analytica Chimica Acta - Volume 594, Issue 1, 26 June 2007, Pages 52–60