Modified ion source triple quadrupole mass spectrometer gas chromatograph for polycyclic aromatic hydrocarbon analyses
ترجمه فارسی عنوان
کروماتوگراف گاز کروماتوگرافی گاز طیف سنج جرم چهارگانه سه فاز اصلاح شده برای تجزیه و تحلیل هیدروکربن های آروماتیک چند حلقه ای
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
چکیده انگلیسی


• We developed a H2-injected, self-cleaning ion source and modified extractor lens.
• 62 polycyclic aromatic hydrocarbons, including MW 302 PAHs, are quantified.
• Previously lumped PAHs – such as chrysene and triphenylene – are separated.
• Sensitivity, accuracy and precision of PAH determinations are exceptional.
• 20 alkylated PAHs are identified, allowing PAH source determinations.

We describe modified gas chromatography electron-impact/triple-quadrupole mass spectrometry (GC–EI/MS/MS) utilizing a newly developed hydrogen-injected self-cleaning ion source and modified 9 mm extractor lens. This instrument, with optimized parameters, achieves quantitative separation of 62 polycyclic aromatic hydrocarbons (PAHs). Existing methods historically limited rigorous identification and quantification to a small subset, such as the 16 PAHs the US EPA has defined as priority pollutants. Without the critical source and extractor lens modifications, the off-the-shelf GC–EI/MS/MS system was unsuitable for complex PAH analysis. Separations were enhanced by increased gas flow, a complex GC temperature profile incorporating multiple isothermal periods, specific ramp rates, and a PAH-optimized column. Typical determinations with our refined GC–EI/MS/MS have a large linear range of 1–10,000 pg μl−1 and detection limits of <2 pg μl−1. Included in the 62 PAHs, multiple-reaction-monitoring (MRM) mode enabled GC-EI/MS/MS identification and quantitation of several constituents of the MW 302 PAH isomers. Using calibration standards, values determined were within 5% of true values over many months. Standard curve r2 values were typically >0.998, exceptional for compounds which are archetypally difficult. With this method benzo[a]fluorene, benzo[b]fluorene, benzo[c]fluorene were fully separated as was benzo[b]fluoranthene, benzo[k]fluoranthene, and benzo[j]fluoranthene. Chrysene and triphenylene, were sufficiently separated to allow accurate quantitation. Mean limits of detection (LODs) across all PAHs were 1.02 ± 0.84 pg μl−1 with indeno[1,2,3-c,d] pyrene having the lowest LOD at 0.26 pg μl−1 and only two analytes above 2.0 pg μl−1; acenaphthalene (2.33 pg μl−1) and dibenzo[a,e]pyrene (6.44 pg μl−1).