Do column frits contribute to the on-column, flow-induced degradation of macromolecules?
ترجمه فارسی عنوان
آیا میله های ستونی به تخریب ماکرومولکول ها ناشی از جریان در ستون کمک می کنند؟
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
چکیده انگلیسی


• Role of column frits in on-column, flow-induced polymer degradation was studied.
• No degradation if frit pore radius is comparable to column hydraulic radius.
• Results explain previously observed degradation in polymer UHPLC.

Flow-induced, on-column degradation is a major hindrance to the accurate characterization of ultra-high molar mass macromolecules and colloids. This degradation is a direct result of the large shear rates which are generated within the column, which cause chain scission to occur both in the interstitial medium and, it has been postulated, at the packing particle pore boundary. An additional putative source of degradation has been the column frits, though little experimental evidence exists to either support or refute this claim. To this effect, the present experiments examine the role of the frits in the degradation of high molar mass macromolecules. Two narrow dispersity polystyrene standards, the molar mass of which differs by a factor of two, were analyzed on three different size-exclusion chromatography (SEC) columns, each with frits of different pore size, at various flow rates. In the smallest pore size column, which also contained the smallest frits and which was packed with the smallest diameter particles, the larger standard was forced to degrade by increasing the flow rate of the mobile phase. During the course of the latter portion of the study, the inlet and the outlet frits were removed from the column, in stepwise fashion. It was concluded that neither frit played any appreciable role in the degradation. Results of our studies were applied to explain previously observed degradation in ultra-high pressure liquid chromatography of polymers. The general conclusion arrived at herein is that the column frits are likely to have a secondary role (as compared to interstitial and pore boundary stresses), or no role at all, in polymer degradation for cases where the frit radius is larger than or equal to the hydraulic radius rcof the column.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography A - Volume 1359, 12 September 2014, Pages 147–155
نویسندگان
,