On the accuracy of gamma spectrometric isotope ratio measurements of uranium
ترجمه فارسی عنوان
بر روی دقت اندازه گیری ایزوتوپهای گاما طیف سنجی اورانیوم
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
چکیده انگلیسی

The isotopic composition of uranium was measured using high resolution gamma spectrometry. Two acid solutions and two samples in the form of UO2 pellets were measured. The measurements were done in close geometries, i.e. directly on the endcap of the high purity germanium detector (HPGe). Applying no corrections for count losses due to true coincidence summing (TCS) resulted in up to about 40% deviation in the abundance of 235U from the results obtained with mass spectrometry. However, after correction for TCS, excellent agreement was achieved between the results obtained using two different measurement methods, or a certified value. Moreover, after corrections, the fitted relative response curves correlated excellently with simulated responses, for the different geometries, of the HPGe detector.