The applicability of LaBr3:Ce detectors for high count rate γ-rayγ-ray spectroscopy is investigated. A 3 in.×3 in. LaBr3:Ce detector is used in a test setup with radioactive sources to study the dependence of energy resolution and photo peak efficiency on the overall count rate in the detector. Digitized traces were recorded using a 500 MHz FADC and analysed with digital signal processing methods. Good performance is obtained using standard techniques up to about 500 kHz counting rate. A pile-up correction method is applied to the data in order to further improve the capabilities at even higher rates with a focus on recovering the losses in efficiency due to signal pile-up. It is shown that γ-rayγ-ray spectroscopy can be performed with only moderate lossen in efficiency and high resolution at count rates even above 1 MHz and that the performance can be enhanced in the region between 500 kHz and 10 MHz by using the applied pile-up correction techniques.
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 686, 11 September 2012, Pages 1–6