A resolution gauge was fabricated for measuring the spatial resolution of radiographic imaging systems. Silicon wafers, 100 mm 〈1 0 0〉, were patterned using standard contact lithography and the patterned features were etched using deep reactive ion etching (DRIE). The smallest features were 5 μm wide line pairs. The resulting etched trenches included aspect ratios of up to 10:1 and were 40.6±0.2 μm deep. The etch rate was 2.7 μm/min for the reported etch depth. A Gd2O3 nanopowder was dispersed into a slurry using Darvan C and water as the dispersant and solvent, respectively. A rapid infiltration method was used to fill the etched trenches with the slurry. Neutron and X-ray radiographs of the resolution gauge and the results demonstrate that the prototype gauge would be a suitable standard for measuring the spatial resolution of both X-ray and neutron radiography systems.
► We describe a method for producing radiographic image spatial resolution gauges.
► Gauge features sizes range from 2 mm to 5 μm.
► The production method yields deep features with nearly perfectly square edges.
► Gadolinium oxide nanopowder was the radiation attenuating medium.
► Results show that the device is suitable for X-ray and neutron radiography systems.
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 646, Issue 1, 1 August 2011, Pages 135–141