Optimisation of the MC-model of a p-type Ge-spectrometer for the purpose of efficiency determination
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Optimisation of the MC-model of a p-type Ge-spectrometer for the purpose of efficiency determination
چکیده انگلیسی

An optimisation of the geometrical model of a p-type detector used for material screening was carried out to improve the accuracy of Monte Carlo simulations in reproducing spectrometric measurements. Gamma-ray sources were measured to determine the dimensions of the detector dead layer and borehole. An agreement between simulations and measurement within 3% was achieved at energies above 100 keV. In contrast, discrepancies on the order of 23% were encountered using the nominal parameters from the detector manufacturer.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 67, Issue 5, May 2009, Pages 706–710