کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1877059 1042042 2009 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
How to assess the sensitivity and related uncertainty of a new solid state passive thoron measuring device
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
How to assess the sensitivity and related uncertainty of a new solid state passive thoron measuring device
چکیده انگلیسی
To test the sensitivity of our CR-39 alpha-track detector thoron passive device (ATD-TnPD), a closed-circuit exposure facility, composed by a new thoron source, a small exposure chamber and a new absolute 220Rn active monitor (TnM), was set up. 220Rn exposure was evaluated by assessing the TnM response and its combined uncertainty. The ATD-TnPD's sensitivity was determined by three thoron exposures, tested at different etching times. A non-linear relationship on etching conditions was obtained, with values in the range 1.54-1.76 (tracks cm2)(kBq h m−3)−1.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 67, Issue 5, May 2009, Pages 854-859