Dose evaluation of an NIPAM polymer gel dosimeter using gamma index
ترجمه فارسی عنوان
ارزیابی دوز از یک پلیمر ژل ناپام با استفاده از شاخص گاما
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
چکیده انگلیسی
Three batches of replicated gel phantoms were independently measured. The average uncertainty of the measurements was less than 1%. The gel was found to have a high degree of spatial uniformity throughout the dosimeter and good temporal stability. A comparison of the line profiles of the treatment planning system and of the data measured by optical CT showed that the dose was overestimated in the penumbra region because of two factors. The first is light scattering due to changes in the refractive index at the edge of the irradiated field. The second is the edge enhancement caused by free radical diffusion. However, the effect of edge enhancement on the NIPAM gel dosimeter is not as significant as that on the BANG gel dosimeter. Moreover, the dose uncertainty is affected by the inaccuracy of the gel container positioning process. To reduce the uncertainty of 3D dose distribution, improvements in the gel container holder must be developed.