BackgroundThe objective of this intravascular ultrasound (IVUS) analysis was to evaluate the vascular response of the thin-strut TAXUS Liberté stent compared with the otherwise identical TAXUS Express stent.Methods and MaterialsThe TAXUS ATLAS and TAXUS ATLAS Long Lesion studies are nonrandomized trials comparing the thin-strut TAXUS Liberté stent to historical TAXUS Express controls from the TAXUS IV and TAXUS V trials. A total of 377 patients enrolled in the two TAXUS ATLAS studies were randomly selected for the IVUS subset and compared to 314 TAXUS Express IVUS controls.ResultsDespite increased lesion complexity in the TAXUS Liberté group, neointimal formation at 9 months was similar in both stents (TAXUS Liberté 13.8±11.0%; TAXUS Express 13.1±13.8%, P=.56). However, this neointima covered more of the overall stent in the TAXUS Liberté (67.9±32.5%) compared with the TAXUS Express (54.4±37.2%, P<.001), suggesting more uniform neointimal distribution. TAXUS Liberté also showed less pronounced negative remodeling at both stent edges and had significantly less (4.3% vs. 9.6%, P=.04) late incomplete stent apposition (ISA).ConclusionsDespite identical polymer and drug release characteristic, the thin-strut TAXUS Liberté stent demonstrates improved neointimal coverage, better edge remodeling, and less late ISA vs. TAXUS Express, hereby highlighting the importance of the platform design for drug-eluting stents.
Journal: Cardiovascular Revascularization Medicine - Volume 12, Issue 4, July–August 2011, Pages 247–257