The DF-4 is a new defibrillator lead technology. We present two cases of non-physiological transient ventricular over-sensing in patients who underwent implantation of an ICD for secondary prevention. Case 1 had ventricular over-sensing during pacing threshold evaluation post defibrillation testing while Case 2 had the lead integrity alert triggered immediately post discharge with transient over-sensing. No lead-connector issues were found. Case 1 was likely due to improper venting of the header and trapped air. Case 2 was hypothesized to be due to intermittent header pin non-contact secondary to blood in the header. These cases reveal that DF-4 leads are subject to both reported and potentially novel causes of transient acute ventricular over-sensing.
Journal: Indian Pacing and Electrophysiology Journal - Volume 15, Issue 4, July–August 2015, Pages 216–219