High performance liquid chromatography with photo diode array for separation and analysis of naproxen and esomeprazole in presence of their chiral impurities: Enantiomeric purity determination in tablets
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
High performance liquid chromatography with photo diode array for separation and analysis of naproxen and esomeprazole in presence of their chiral impurities: Enantiomeric purity determination in tablets
چکیده انگلیسی


- HPLC-DAD for enantioseparation of naproxen and esomeprazole four isomers.
- Chiral analysis of S-isomers with high selectivity and sensitivity.
- High resolution values for naproxen and omeprazole four isomers in presence of IS.
- Enantiomeric purity was applied to S-naproxen and esomeprazole tablets.
- Chiral impurities (R-isomers) detected at 1% level.

A stereoselective high performance liquid chromatographic method with diode array detection (HPLC-DAD) was introduced for S-naproxen and esomeprazole determination in tablets. The separation was achieved on a Kromasil Cellucoat chiral column using a mobile phase consisting of hexane: isopropanol: trifluoroacetic acid (TFA) (90:9.9:0.1 v/v/v). The proposed system was found to be suitable for the enantioseparation of naproxen and omeprazole biologically active isomers. After optimization of the chromatographic conditions, resolution values of 3.84 and 2.17 could be obtained for naproxen and omeprazole isomers, respectively. The method was fully validated for the determination of S-isomers of each drug in their dosage form. Also, the enentiomeric purity was determined in commercial tablet containing S-naproxen and esomeprazole. The enantiomeric purity was calculated for each drug and the chiral impurities (R-isomers) could be determined at 1% level. The method was validated and good results with respect to linearity, precision, accuracy, selectivity and robustness were obtained. The limits of detection (LOD) and quantification (LOQ) were 2.00, 6.50 and 0.10, 0.35 μg mL−1 for S-naproxen and esomeprazole, respectively.