کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8249774 1533376 2018 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Experimental study on fluence rate response of LaBr3 to pulsed X-rays
ترجمه فارسی عنوان
بررسی تجربی بر پاسخ فرکانسی لانگر 3 به اشعه ایکس پالسی
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
چکیده انگلیسی
The linear response is the key property employed by a radiation detector to determine the accurate intensity of a pulse radiation field. In this study, we investigated the high fluence rate response behavior of a LaBr3 scintillator. We used a high-intensity pulsed X-ray source called “QiangGuang-I” which could produce an X-ray pulse with a total dose of 100 Gy, full width at half maximum of 20 ± 5 ns, and an average energy of 1 MeV, to test the linear response of the LaBr3 scintillator. The Lissajous figure method was used in this experiment. The results showed that the fluence rate linear response limit was more than 1.8 × 1020 MeV/(cm2·s).
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Measurements - Volume 114, July 2018, Pages 8-11