کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
870505 909872 2005 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Glutamate detection from nerve cells using a planar electrodes array integrated in a microtiter plate
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Glutamate detection from nerve cells using a planar electrodes array integrated in a microtiter plate
چکیده انگلیسی

There is an increasing interest in new strategies for replacing animal tests in research. The use of cell cultures and integrated electrodes is seen as a promising alternative that could potentially solve this problem. In this work, we present a l-glutamate sensor based on a bienzyme redox hydrogel, capable of detecting the release of this excitatory neurotransmitter from adherently growing cells upon stimulation. The low working potential required for the operation of the sensor decreases the possibility of interference by easily oxidizable compounds always present in complex biological samples. A low detection limit of 0.5 μM l-glutamate, a response time of about 35 s, and a linear range of up to 60 μM are the main characteristics of the sensor. The system has been successfully employed to monitor the release of l-glutamate from HN10 and C6 cells upon stimulation with K+-ions. The developed integrated electrochemical platform will be used in future for drug screening and potentially for replacing animal models in neurological experiments.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biosensors and Bioelectronics - Volume 20, Issue 10, 15 April 2005, Pages 2116–2119
نویسندگان
, , , , ,