• Size exclusion separation of polysaccharide with reverse phase column.
• Use the entire separation window before and after void volume to tackle complex separation problem.
• Combined use of size and hydrophobicity based separation in RPLC.
This work describes the use of a reverse phase (RP) column for size-based separation of hydrophilic polysaccharides. The entire separation window (before and after void volume) is used to provide unique separation of polysaccharides from other components in a glycol-conjugate vaccine or complex media of fermentation broth. The technique has also been applied to the separation of polysaccharides of different sizes. The effect of chromatographic parameters including type of packing material (CN, C8 and C18), pore size (80 and 300 Å) of stationary phase, concentration of organic solvent on the separation was investigated. In addition, characterization of size-based separation was evaluated by coupling of RP column with multi-angle laser light scattering (MALLs) detector.
Journal: Journal of Chromatography A - Volume 1323, 3 January 2014, Pages 97–103