کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1206718 | 1493742 | 2007 | 9 صفحه PDF | دانلود رایگان |
Mobile phase compensation, first reported for the charged aerosol detector (CAD), was used as a suitable method to overcome problems related to the mobile phase-dependent response of the evaporative light scattering detector (ELSD). Mobile phase compensation was effectively performed both in the flow injection- and in gradient modes. Without compensation, the response factors of the ELSD for six sulfonamide drugs differed by a factor of two when varying the mobile phase composition between 10 and 90% acetonitrile. This change could be effectively eliminated using the technique of mobile phase compensation, where a secondary pump with a reversed gradient was used to provide the detector with a constant composition of the mobile phase. For identical experimental conditions, the ELSD showed a nearly constant, albeit somewhat reduced, response with compensation. This indicates that under such conditions, the ELSD behaved as a concentration-sensitive detector. The analysis of sulfonamides drugs at 0.05% level using gradient UPLC-ELSD separation with mobile phase compensation is demonstrated.
Journal: Journal of Chromatography A - Volume 1161, Issues 1–2, 17 August 2007, Pages 183–191