کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1216047 1494160 2009 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A highly sensitive and robust UPLC-MS with electrospray ionization method for quantitation of taxifolin in rat plasma
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
A highly sensitive and robust UPLC-MS with electrospray ionization method for quantitation of taxifolin in rat plasma
چکیده انگلیسی
A sensitive ultra performance liquid chromatography-mass spectrometry method has been developed and validated for the quantification of taxifolin in rat plasma. Following liquid/liquid extraction by ethyl acetate, the analytes were separated on a Sunfire™ (2.1 mm × 50 mm, 3.5 μm) column and analyzed in the selected ion recording with a negative electrospray ionization mode. The method was linear over the concentration range of 6-6750 ng/mL. Intra- and inter-day precisions were all within 8% and accuracy ranged from 92.9% to 105.1%. The lower limit of quantification was 6 ng/mL. The present method was successfully applied to the estimation of the pharmacokinetic parameters of taxifolin following intravenous and oral administration to rats. The absolute bioavailability of taxifolin was 0.17% in rat.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography B - Volume 877, Issues 18–19, 15 June 2009, Pages 1778-1786
نویسندگان
, , , , , ,