کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2697402 1143906 2006 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Relative performance of soft contact lenses having lathe-cut posterior surfaces with and without additional polishing
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی چشم پزشکی
پیش نمایش صفحه اول مقاله
Relative performance of soft contact lenses having lathe-cut posterior surfaces with and without additional polishing
چکیده انگلیسی

After a preliminary investigation of the effects of tool feed rate and spindle speed on the surface roughness of unhydrated, lathe-cut polymacon surfaces, a laboratory and clinical comparison was made between lenses with identical parameters except that the lathe-cut posterior surface was left unpolished in the “test” lenses and was polished in the “control” lenses. The lenses had moulded anterior surfaces. Laboratory comparisons included surface roughness, lens power and its uniformity across the surface. Double-blind clinical trials over 4-hour (27 subjects) and 1-month (10 subjects) periods, involved one eye of each subject wearing a “test” lens and the other, a “control” lens. No clinically significant differences were found between the results for the test and control lenses. It is concluded that today's lathing technology makes a final polishing stage unnecessary.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Contact Lens and Anterior Eye - Volume 29, Issue 2, May 2006, Pages 101–107
نویسندگان
, ,