کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2928479 1176211 2014 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Inappropriate Detection of a Supraventricular Tachycardia as Dual Tachycardia by the PR Logic™ Algorithm
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Inappropriate Detection of a Supraventricular Tachycardia as Dual Tachycardia by the PR Logic™ Algorithm
چکیده انگلیسی
Tachycardia detection and therapy algorithms in Implantable Cardioverter-Defibrillators (ICD) reduce, but do not eliminate inappropriate ICD shocks. Awareness of the pros and cons of a particular algorithm helps to predict its utility in specific situations. We report a case where PR logic™, an algorithm commonly used in currently implanted ICDs to differentiate supraventricular tachycardia (SVT) from ventricular tachycardia resulted in inappropriate detection and shock for an SVT, and discuss several solutions to the problem.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Indian Pacing and Electrophysiology Journal - Volume 14, Issue 3, May–June 2014, Pages 161-164
نویسندگان
, , , , , , , ,