Ultrasound–fluoroscopy registration for prostate brachytherapy dosimetry
موضوعات مرتبط
مهندسی و علوم پایه مهندسی کامپیوتر گرافیک کامپیوتری و طراحی به کمک کامپیوتر
پیش نمایش صفحه اول مقاله
Ultrasound–fluoroscopy registration for prostate brachytherapy dosimetry
چکیده انگلیسی

Prostate brachytherapy is a treatment for prostate cancer using radioactive seeds that are permanently implanted in the prostate. The treatment success depends on adequate coverage of the target gland with a therapeutic dose, while sparing the surrounding tissue. Since seed implantation is performed under transrectal ultrasound (TRUS) imaging, intraoperative localization of the seeds in ultrasound can provide physicians with dynamic dose assessment and plan modification. However, since all the seeds cannot be seen in the ultrasound images, registration between ultrasound and fluoroscopy is a practical solution for intraoperative dosimetry. In this manuscript, we introduce a new image-based nonrigid registration method that obviates the need for manual seed segmentation in TRUS images and compensates for the prostate displacement and deformation due to TRUS probe pressure. First, we filter the ultrasound images for subsequent registration using thresholding and Gaussian blurring. Second, a computationally efficient point-to-volume similarity metric, an affine transformation and an evolutionary optimizer are used in the registration loop. A phantom study showed final registration errors of 0.84 ± 0.45 mm compared to ground truth. In a study on data from 10 patients, the registration algorithm showed overall seed-to-seed errors of 1.7 ± 1.0 mm and 1.5 ± 0.9 mm for rigid and nonrigid registration methods, respectively, performed in approximately 30 s per patient.

Figure optionsDownload high-quality image (257 K)Download as PowerPoint slideHighlights
► A nonrigid intensity-based ultrasound–fluoroscopy registration method for prostate brachytherapy dosimetry is introduced.
► The affine registration compensates for the probe pressure effects on the prostate.
► The registration algorithm obviates the need for manual seed segmentation in TRUS.
► Seed-to-seed registration errors of 1.5 ± 0.9 mm are achieved.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Medical Image Analysis - Volume 16, Issue 7, October 2012, Pages 1347–1358
نویسندگان
, , , , , , , , , ,