Gas chromatographic quadrupole time-of-flight full scan high resolution mass spectrometric screening of human urine in antidoping analysis
چکیده انگلیسی


- Steroid profile obtained with high-resolution GC/QTOF meet the WADA specifications.
- Specificity of the current method is similar to the routine method based on tandem MS in a GC/QQQ.
- The acquisition FS data of GC/QTOF allow the retrospective analysis to detect unknown substances.

This paper presents the development and validation of a high-resolution full scan (FS) electron impact ionization (EI) gas chromatography coupled to quadrupole Time-of-Flight mass spectrometry (GC/QTOF) platform for screening anabolic androgenic steroids (AAS) in human urine samples. The World Antidoping Agency (WADA) enlists AAS as prohibited doping agents in sports, and our method has been developed to comply with the qualitative specifications of WADA to be applied for the detection of sports antidoping prohibited substances, mainly for AAS. The method also comprises of the quantitative analysis of the WADA's Athlete Biological Passport (ABP) endogenous steroidal parameters. The applied preparation of urine samples includes enzymatic hydrolysis for the cleavage of the Phase II glucuronide conjugates, generic liquid-liquid extraction and trimethylsilyl (TMS) derivatization steps. Tandem mass spectrometry (MS/MS) acquisition was applied on few selected ions to enhance the specificity and sensitivity of GC/TOF signal of few compounds. The full scan high resolution acquisition of analytical signal, for known and unknown TMS derivatives of AAS provides the antidoping system with a new analytical tool for the detection designer drugs and novel metabolites, which prolongs the AAS detection, after electronic data files' reprocessing. The current method is complementary to the respective liquid chromatography coupled to mass spectrometry (LC/MS) methodology widely used to detect prohibited molecules in sport, which cannot be efficiently ionized with atmospheric pressure ionization interface.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography B - Volume 1063, 15 September 2017, Pages 74-83