کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5454456 1514336 2017 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Assessment of patients X-ray doses at three government hospitals in Duhok city lacking requirements of effective quality control
ترجمه فارسی عنوان
ارزیابی بیماران دوزهای اشعه ایکس در سه بیمارستان دولتی در شهر دهوک بدون نیاز به کنترل کیفیت موثر
موضوعات مرتبط
مهندسی و علوم پایه مهندسی مواد دانش مواد (عمومی)
چکیده انگلیسی
s: Significant variations between exposure factors recorded in this study and those recommended in the context of quality criteria and standards were shown. The results have also shown that about twenty percent of the patient doses (ESD and E) were equal or below the recommended values of the diagnostic reference levels (DRLs). For abdomen, pelvis and skull examinations, the ESD values were slightly above the diagnostic reference levels. For chest and cervical the ESD values were much higher than diagnostic reference levels. These values were more reasonable only in one hospital. High ESD values can be attributed to the slightly higher tube voltages and lower mAs values that were used. The high patient dose values suggest that any adequate change of the exposure parameters that aims at the reduction of dose must be done without compromising the image quality. This study recommends a quick action toward implementing a quality control program and employing special staff of medical physicists in the evaluated hospitals.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Radiation Research and Applied Sciences - Volume 10, Issue 3, July 2017, Pages 183-187
نویسندگان
, ,