Reliability of a clinical 3D freehand ultrasound technique: Analyses on healthy and pathological muscles
موضوعات مرتبط
مهندسی و علوم پایه مهندسی کامپیوتر علوم کامپیوتر (عمومی)
پیش نمایش صفحه اول مقاله
Reliability of a clinical 3D freehand ultrasound technique: Analyses on healthy and pathological muscles
چکیده انگلیسی
When only 10% of the US images were outlined, there was an average underestimation of muscle volume of 1.1% and 1.6% with respect the computation of all the available images, for the typically developing and spastic cerebral palsy groups, respectively. For both groups, the reliability was higher for data processing than for data acquisition. High inter-class correlation coefficient values were found for processing and acquisition reliability, with worst case values of 0.89 and 0.61, respectively. The standard error of measurement, expressed as a percentage of the average volumes, was smaller than 2.6 ml (4.8%) in all cases. Conclusions: The present analysis demonstrates the effectiveness of applying 3D freehand ultrasonography in a clinical setting for analysing healthy and pathological paediatric muscle.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Computer Methods and Programs in Biomedicine - Volume 156, March 2018, Pages 97-103