کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10351263 864346 2005 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Effects of monitor luminance change on observer detection performance
موضوعات مرتبط
مهندسی و علوم پایه مهندسی کامپیوتر نرم افزارهای علوم کامپیوتر
پیش نمایش صفحه اول مقاله
Effects of monitor luminance change on observer detection performance
چکیده انگلیسی
To confirm that the dimming luminance change in the CRT monitor will have a detrimental effect on diagnostic performance, we performed the four-alternative forced choice experiments on the CRT monitor with the 11 luminance conditions simulating CRT degraded by long-term usage. Six radiologists and one pulmonary physician performed a detection task for 11,000 test areas on 110 test images. The 11 monitor conditions significantly affected the correct detection rates of simulated nodules in the four-alternative forced choice experiments. The gradual deterioration in CRT monitor luminance will have certainly a detrimental effect on detection performance.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Computerized Medical Imaging and Graphics - Volume 29, Issue 1, January 2005, Pages 35-41
نویسندگان
, , , , , ,