کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10410459 894437 2005 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Electrospun fibrous polyacrylic acid membrane-based gas sensors
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Electrospun fibrous polyacrylic acid membrane-based gas sensors
چکیده انگلیسی
Electrospun fibrous polyacrylic acid (PAA) membranes were studied as sensing material coated on quartz crystal microbalance (QCM) for ammonia detection at ppb level. The fibrous membranes with different morphology can be deposited on the electrode of QCM by electrospinning the PAA solutions with various solvent composition of H2O and ethanol. The results of sensing experiments indicated that the sensitivity of the fibrous membranes coated QCM (FM-QCM) sensors was four times higher than that of the casting films coated QCM (CF-QCM) sensors. Meanwhile, the FM-QCM sensors exhibited high sensitivity towards low concentration of ammonia, as low as 130 ppb at the relative humidity of 40%. The pre-sorbed water in fibrous membranes was proved to be the key factor to affect the sensitivity of FM-QCM sensors for ammonia. The sensor performance has been found to depend on the morphology of the fibrous membranes, concentration of ammonia, coating load of the fibrous membranes on QCM, and relative humidity. Preliminary study of the stability investigation was also presented.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Sensors and Actuators B: Chemical - Volume 106, Issue 1, 29 April 2005, Pages 477-483
نویسندگان
, , ,