A quantitative lateral flow assay to detect complement activation in blood
ترجمه فارسی عنوان
آزمایش کمی جریان جانبی برای تشخیص فعال شدن کمپلکس در خون
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
چکیده انگلیسی

Complement is a major effector arm of the innate immune system that responds rapidly to pathogens or altered self. The central protein of the system, C3, participates in an amplification loop that can lead to rapid complement deposition on a target and, if excessive, can result in host tissue damage. Currently, complement activation is routinely monitored by assessing total C3 levels, which is an indirect and relatively insensitive method. An alternative approach would be to measure downstream C3 activation products such as C3a and iC3b. However, in vitro activation can produce falsely elevated levels of these biomarkers. To circumvent this issue, a lateral flow immunoassay system was developed that measures iC3b in whole blood, plasma, and serum and avoids in vitro activation by minimizing sample handling. This assay system returns results within 15 min and specifically measures iC3b while having minimal cross-reactivity to other C3 split products. While evaluating the potential of this assay, it was observed that circulating iC3b levels can distinguish healthy individuals from those with complement activation-associated diseases. This tool is engineered to provide an improved method to assess complement activation at point of care and could facilitate studies to monitor disease progression in a variety of inflammatory conditions.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Analytical Biochemistry - Volume 477, 15 May 2015, Pages 78–85
نویسندگان
, , , , , , , , , , ,