Calibration in isothermal titration calorimetry: Heat and cell volume from heat of dilution of NaCl(aq)
چکیده انگلیسی

An isothermal titration calorimeter of the perfusion type (MicroCal model VP-ITC) is calibrated using the heat of dilution of NaCl in water. The relative apparent molar enthalpy function (Lϕ) for NaCl(aq) varies strongly and nonlinearly with concentration in the low-concentration region (<0.2 M) that is sampled easily and extensively in a single program of injections of NaCl solution into water. This nonlinearity makes it possible to calibrate with respect to two quantities: the measured heat and the active cell volume. The heat factor is determined with typical standard error 0.003; its value in the current case is 0.987. The cell volume factor is 0.93 but is quite sensitive to possible systematic errors in the temperature and in the literature values for Lϕ. Both correction factors are closely tied to the delivered volume from the injection syringe, which required a correction factor of 0.973, attributed to an instrumental gear ratio error. Temperature calibration of the instrument showed a small offset of 0.12 K at the temperature 25 °C of the experiments, but the error increased to more than 1 K at 46 °C. The experiments were not able to distinguish clearly between mixing algorithms that assume instantaneous mixing on injection and those that assume instantaneous injection followed by mixing; however, examination of these algorithms has revealed an error in a program widely used to analyze isothermal titration calorimetry data.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Analytical Biochemistry - Volume 360, Issue 1, 1 January 2007, Pages 47–55
نویسندگان
,