Octupolar excitation has been investigated with respect to mass-selective ion centering in buffer-gas filled Penning traps for the isolation of an ion species of interest. To this end, off-line test measurements have been performed with the preparation Penning trap of the mass spectrometer ISOLTRAP located at ISOLDE/CERN including a comparison with the conventional quadrupolar excitation. The experimental results are compared to numerical simulations of ions under influence of quadrupolar and octupolar excitation.
Figure optionsDownload high-quality image (184 K)Download as PowerPoint slideHighlights
► We investigate the applicability of octupolar excitation in buffer-gas filled Penning traps.
► The experiment yields a drop of resolving power for octupolar excitation as compared to the expectations.
► A simulation including ion-atom collisions has been written in order to explain the measurements.
► The simulation shows a crucial influence of the buffer-gas on the ion trajectory during octupolar excitation.
Journal: International Journal of Mass Spectrometry - Volume 314, 15 March 2012, Pages 6–12