• Abundance sensitivity of mass spectrometers was discussed extensively.
• Abundance sensitivity could be improved by adding a quadrupole to magnetic sector.
• Tandem mass spectrometer of magnetic-electric-quadrupole type was developed.
• Abundance sensitivity of better than 6 × 10−10 was achieved.
• Improving scheme was proposed for better sensitivity.
A novel magnetic-electric-quadrupole (MEQ) tandem mass spectrometer is introduced in this paper. As there is a huge gap between the kinetic energies of ions in the magnetic sector and that in the quadrupole mass spectrometers, no sector-quadrupole type tandem mass spectrometer has been developed before. However, with the help of a set of specially designed ion-optical lens, we have developed an MEQ tandem mass spectrometer. Measurement result shows that the abundance sensitivity of this instrument was better than 6 × 10−10, which was nearly 3 orders of magnitude higher than the tandem mass spectrometer without the quadrupole.
Figure optionsDownload high-quality image (220 K)Download as PowerPoint slide
Journal: International Journal of Mass Spectrometry - Volumes 343–344, 1 June 2013, Pages 9–14