کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1228023 | 968443 | 2011 | 10 صفحه PDF | دانلود رایگان |

This paper explores the potential of applying reversed-phase sequential injection chromatography (SIC) to determine the anticoccidial agents Lasalocid and Toltrazuril in various matrices including ground water, pharmaceutical formulations and feed. SIC was performed by connecting a 25 × 4.6 mm monolithic C18 column to a 2 m long pathlength capillary flow cell, where the usage of a flow cell lowers the detection limit compared to a conventional short-distance flow cell, providing a simple detection system for these two compounds which are initially poorly UV absorbents.The proposed set-up provides a high injection throughput of 12 h− 1, as well as a limit of detection of 0.019 and 0.010 mg/L for Toltrazuril and Lasalocid, respectively. The repeatabilities obtained (n = 10) were lower than 2% and 4% for Toltrazuril and Lasalocid, respectively.
Journal: Microchemical Journal - Volume 98, Issue 2, July 2011, Pages 190–199