کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1228384 968464 2007 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Determination of trace levels of mercury in aqueous solutions by inductively coupled plasma atomic emission spectrometry: Elimination of the ‘memory effect’
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Determination of trace levels of mercury in aqueous solutions by inductively coupled plasma atomic emission spectrometry: Elimination of the ‘memory effect’
چکیده انگلیسی

The measurement of mercury in aqueous solutions by ICP-AES is adversely affected by the memory effect wherein mercury accumulates within the sample introduction system and is slowly released over time to give increasing response signals at the same initial mercury concentration. The memory effect is obviated by the addition of Hg(II) complexants: thiourea and gold(III) chloride are both effective in preventing mercury sorption and vapor buildup with the latter being preferred because the memory effect vanishes more rapidly. Conditions are described wherein it is possible to quantify low levels of mercury(II) in aqueous solutions by ICP-AES under routine operating conditions that can be applied to other metal ions by adding 1 mg of gold(III) chloride per 3 mg of mercury(II) to those solutions.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Microchemical Journal - Volume 86, Issue 1, June 2007, Pages 37–41
نویسندگان
, ,