کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1252897 1496057 2011 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Microchip based sample treatment device interfaced with ICP-MS for the analysis of transition metals from environmental samples
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی (عمومی)
پیش نمایش صفحه اول مقاله
Microchip based sample treatment device interfaced with ICP-MS for the analysis of transition metals from environmental samples
چکیده انگلیسی

A downscaled solid phase (SPE) device applicable for sample preparation prior to ICP-MS monitoring, have been constructed making use of the lab on a chip concept. Standard photolithography and wet chemical etching were used to fabricate glass microfluidic devices accommodating three microchannels, each of them incorporating a defined section that could be packed with SPE materials; selective chelating resin. The microfluidic device was interfaced with the ICP-MS instrument throughout a low flow rate concentric nebuliser using a Teflon connector, and coupled with a flow injection manifold delivering samples and reagents via a manually operated splitting valve. The feasibility of the miniaturized prototype to perform SPE of trace metals was proved by analyzing trace metals, Cd, Co, and Ni, in seawater reference materials; CASS-2 and SLEW-1. The obtained result was in good agreement with the certified values. The device could be used as a remote miniaturized sample treatment for field work.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Arabian Journal of Chemistry - Volume 4, Issue 2, April 2011, Pages 195–203
نویسندگان
, ,