The present work deals with the design of a multi-probe support for simultaneous multi-point measurements. The article analyses the effects of the insertion of a hot-wire rake in a test section where the flow shows not constant temperatures. The constant temperature anemometry technique is known to have a relevant sensibility to both fluid temperature variations during the measurements and temperature differences between calibration and testing conditions. Therefore a technique to take into account for the flow temperature drifts influence is proposed and validated. The temperature correction presented allows reducing the influence of temperature variations on the measured velocity. This is achieved introducing a correction term, namely the temperature-loading factor that can be optimized for the individual probe and measurement conditions. The correction also takes into account for variations in fluid property values (Prandtl number, dynamic viscosity, heat conductivity and density) with temperature.
Journal: Energy Procedia - Volume 85, January 2016, Pages 35-43