• The cross-talk in a row–column addressed 2-D CMUT array was investigated.
• The array was used as both a 1-D array and as a 2-D row–column array.
• The cross-talk was 24 dB lower in the 2-D row–column setup than in the 1-D setup.
• The cross-talk was predicted by theoretical dispersion relations.
• The dominating cross-talk mechanism was guided interface modes.
The acoustical cross-talk in row–column addressed 2-D transducer arrays for volumetric ultrasound imaging is investigated. Experimental results from a 2.7 MHz, λ/2λ/2-pitch capacitive micromachined ultrasonic transducer (CMUT) array with 62 rows and 62 columns are presented and analyzed in the frequency-wavenumber domain. The sources of cross-talk are identified and predicted theoretically. The nearest neighbor cross-talk is -23.9±3.7-23.9±3.7 dB when the array is used as a 1-D array with the rows functioning as both transmitters and receivers. In the row–column configuration, with the columns transmitting and the rows receiving, the cross-talk is reduced to -40.2±3.5-40.2±3.5 dB.
Journal: Ultrasonics - Volume 63, December 2015, Pages 174–178