Influence of the ultrasound transducer bandwidth on selection of the complementary Golay bit code length
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم آکوستیک و فرا صوت
پیش نمایش صفحه اول مقاله
Influence of the ultrasound transducer bandwidth on selection of the complementary Golay bit code length
چکیده انگلیسی
In contrast to previously published papers [A. Nowicki, Z. Klimonda, M. Lewandowski, J. Litniewski, P.A. Lewin, I. Trots, Comparison of sound fields generated by different coded excitations - Experimental results, Ultrasonics 44 (1) (2006) 121-129; J. Litniewski, A. Nowicki, Z. Klimonda, M. Lewandowski, Sound fields for coded excitations in water and tissue: experimental approach, Ultrasound Med. Biol. 33 (4) (2007) 601-607], which examined the factors influencing the spatial resolution of coded complementary Golay sequences (CGS), this paper investigates the effect of ultrasound imaging transducer's fractional bandwidth on the gain of the compressed echo signal for different spectral widths of the CGS. Two different bit lengths were considered, specifically one and two cycles. Three transducers having fractional bandwidth of 25%, 58% and 80% and operating at frequencies 6, 4.4 and 6 MHz, respectively were examined (one of the 6 MHz sources was focused and made of composite material). The experimental results have shown that by increasing the code length, i.e. decreasing the bandwidth, the compressed echo amplitude could be enhanced. The smaller the bandwidth was the larger was the gain; the pulse-echo sensitivity of the echo amplitude increased by 1.88, 1.62 and 1.47, for 25%, 58% and 80% bandwidths, respectively. These results indicate that two cycles bit length excitation is more suitable for use with bandwidth limited commercially available imaging transducers. Further, the time resolution is retained for transducers with two cycles excitation providing the fractional bandwidth is lower than approximately 90%. The results of this work also show that adjusting the code length allows signal-to-noise-ratio (SNR) to be enhanced while using limited (less that 80%) bandwidth imaging transducers. Also, for such bandwidth limited transducers two cycles excitation would not decrease the time resolution, obtained with “conventional” spike excitation. Hence, CGS excitation could be successfully implemented with the existing, relatively narrow band imaging transducers without the need to use usually more expensive wideband, composite ones.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Ultrasonics - Volume 47, Issues 1–4, December 2007, Pages 64-73
نویسندگان
, , , , ,