چکیده انگلیسی

In order to guide the needle to the correct location in 3D US-guided brachytherapy, the needle is continuously tracked as it is being inserted. A pre-scan before the needle insertion and a post-scan after the needle insertion are subtracted to obtain a difference image containing the needle. The image is projected along two orthogonal directions approximately perpendicular to the needle, and the 3D needle is reconstructed from the segmented needles in the two projected images. The seeds implanted with the needle are located in the cropped volume along the needle. Thus, the seeds are segmented using a tri-bar model and 3D line segment patterns. Finally, the positions of the seeds are determined using a peak detection technique. Experiments with agar and turkey/chicken phantoms as well as patient data demonstrated that our needle segmentation technique could segment the needle in near real-time with an accuracy of 0.6 mm in position and 1.0° in orientation. The true-positive rate for seed segmentation is 100% for the agar phantom and 93% for the chicken phantom. The average distance to manual seed segmentation was 1.0 mm for the agar phantom and 1.7 mm for the chicken phantom.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Ultrasonics - Volume 44, Supplement, 22 December 2006, Pages e331–e336
نویسندگان
, , , , ,