کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
181592 459405 2009 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A sensitivity-controlled hydrogen peroxide sensor based on self-assembled Prussian Blue modified electrode
موضوعات مرتبط
مهندسی و علوم پایه مهندسی شیمی مهندسی شیمی (عمومی)
پیش نمایش صفحه اول مقاله
A sensitivity-controlled hydrogen peroxide sensor based on self-assembled Prussian Blue modified electrode
چکیده انگلیسی

In this communication, a hydrogen peroxide (H2O2) sensor based on self-assembled Prussian Blue (PB) modified electrode was reported. Thin film of PB was deposited on the electrode by self-assembly process including multiple sequential adsorption of ferric ions and hexacyanoferrate ions. The as-prepared PB modified electrode displayed sufficient stability for practical sensing application. At an applied potential of −0.05 V vs. Ag/AgCl (sat. KCl), PB modified electrode with 30 layers exhibited a linear dependence on H2O2 concentration in the range of 1 × 10−6–4 × 10−4 M (r = 0.9998) with a sensitivity of 625 mA M−1 cm−2. It was found that the sensitivity of H2O2 sensors could be well controlled by adjusting the number of deposition cycles for PB preparation. This work demonstrates the feasibility of self-assembled PB modified electrode in sensing application, and provides an effective approach to control the sensitivity of PB-based amperometric biosensors.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Electrochemistry Communications - Volume 11, Issue 2, February 2009, Pages 484–487