Application of Monte Carlo calculations for validation of a treatment planning system in high dose rate brachytherapy
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم فیزیک هسته ای و انرژی بالا
پیش نمایش صفحه اول مقاله
Application of Monte Carlo calculations for validation of a treatment planning system in high dose rate brachytherapy
چکیده انگلیسی

AimThe accuracy of treatment planning systems is of vital importance in treatment outcomes in brachytherapy. In the current study the accuracy of dose calculations of a high dose rate (HDR) brachytherapy treatment planning system (TPS) was validated using the Monte Carlo method.Materials and methodsThree 60Co sources of the GZP6 afterloading brachytherapy system were modelled using MCNP4C Monte Carlo (MC) code. The dose distribution around all the sources was calculated by MC and a dedicated treatment planning system. The results of both methods were compared.ResultsThere was good agreement (<2%) between TPS and MC calculated dose distributions except at a point near the sources (<1 cm) and beyond the tip of the sources.ConclusionsOur study confirmed the accuracy of TPS calculated dose distributions for clinical use in HDR brachytherapy.