Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose
چکیده انگلیسی


• HTR was measured for proton beams with energies ranging from 14.5 to 58 MeV.
• HTR was measured for proton doses ranging from 0.5 to 30 Gy.
• HTR was compared with the relative TL efficiency to proton beams.

The high-temperature ratio (HTR) is a parameter quantifying changes of the shape of the high-temperature part of the LiF:Mg,Ti glow-curve after exposure to densely ionizing radiation. It was introduced in order to estimate the ‘effective LET’ of an unknown radiation field and to correct the decreased relative TL efficiency for high-Linear Energy Transfer (LET) radiation.In the present work the dependence of HTR on proton energy (14.5–58 MeV) and dose (0.5–30 Gy) was investigated. All measured HTR values were at the level of 1.2 or higher, therefore significantly different from the respective value for gamma-rays (HTR = 1), but HTR was found to be insensitive to changes of proton energy above 20 MeV. As a result the relationship between HTR and relative TL efficiency is not unequivocal. The HTR was found to be dependent on absorbed dose even for the lowest studied doses.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Measurements - Volume 71, December 2014, Pages 39–42