PADC detectors are commonly used both as radon detectors and as personal dosemeters for neutron radiation. In both cases the measurement in workplace fields are characterized by long term irradiations, lasting up to several months. During this period the detectors undergo to uncontrolled and unmonitored environmental conditions. In a recent work the same authors demonstrated that the environmental conditions, mainly temperature, can seriously affect the detector sensitivity. The temperature effect is to reduce the V ratio by decreasing the track etching velocity Vt. This causes a decrease in limit angle and efficiency.This paper describes the same ageing/fading test made with CR39 from a different supplier. In this case it seems that the temperature has a negligible effect, so no compensation algorithm is needed.
► One year measurement to evaluate the ageing and fading effects on PADC detectors.
► Comparison of detectors from different suppliers.
► Measurement of the detectors efficiency as a function of ageing and fading.
Journal: Radiation Measurements - Volume 46, Issue 10, October 2011, Pages 1160–1167