کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1881683 | 1043169 | 2010 | 5 صفحه PDF | دانلود رایگان |
عنوان انگلیسی مقاله ISI
Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
موضوعات مرتبط
مهندسی و علوم پایه
فیزیک و نجوم
تشعشع
پیش نمایش صفحه اول مقاله

چکیده انگلیسی
In this work the feasibility of using Mg2SiO4:Tb as a fiberoptic radioluminescent (RL) dosimetry probe for real-time dosimetry has been investigated for the first time. In particular, the stability of the RL signal after repeated use, the spectrum of the RL emission and the dose-rate response curve of a Mg2SiO4:Tb-based fiberoptic probe have been determined. The probe has been also used to obtain a percentage dose depth curve in a water phantom and its performance has been compared to that of a standard ion chamber. Besides, its absolute RL yield has been compared to that of an RL probe based on the commercial Al2O3:C phosphor.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Measurements - Volume 45, Issue 1, January 2010, Pages 78–82
Journal: Radiation Measurements - Volume 45, Issue 1, January 2010, Pages 78–82
نویسندگان
P. Molina, M. Prokic, J. Marcazzó, M. Santiago,