کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1881740 1533444 2009 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Measurements of fast deuterons from plasma accelerator by means of PM-355 track detectors
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Measurements of fast deuterons from plasma accelerator by means of PM-355 track detectors
چکیده انگلیسی
This paper reports on studies of fast ions (mostly deuterons) emitted from an RPI (Rod-Plasma-Injector) plasma accelerator. The first aim was the verification of applicability of PM-355 track detectors for dosimetry of lower-energy (<200 KeV) deuterons. The second aim was information about energy spectra of deuterons from RPI-IBIS device in different operational modes. Mass- and energy spectra of ions were investigated with a Thomson-analyzer and PM-355 detectors. On the recorded deuteron-parabolas we chose points, and from numbers of tracks we determined the deuteron energy distribution. In the slow-mode the energy distribution peak was observed at about 40 keV, while the maximum energy amounted to about 150 keV. Those measurements confirmed an influence of the initial gas-conditions on energy distributions of the deuteron streams. The results are of importance for plasma-physics and applications. Another result was the confirmation that PM-355 detectors might be used for accurate measurements of low-energy (<200 keV) deuterons.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Measurements - Volume 44, Issues 9–10, October–November 2009, Pages 870-873
نویسندگان
, , , ,